beautypg.com

Operational theory – Teledyne 3000ZA - Trace oxygen analyzer User Manual

Page 20

background image

Operational Theory

3000ZA

Teledyne Analytical Instruments

8

Operational Theory

2.1 Introduction

The analyzer is composed of three subsystems:

1. Zirconium Oxide (ZrO

2

) Sensor

2. Sample System
3. Electronic Signal Processing, Display and Control

The sample system is designed to accept the sample gas and

transport it through the analyzer without contaminating or altering the
sample prior to analysis. The ZrO

2

sensor is a device that translates the

amount of oxygen present in the sample into an electrical current. The
electronic signal processing, display and control subsystem simplifies
operation of the analyzer and accurately processes the sampled data. The
microprocessor controls all signal processing, input/output and display
functions for the analyzer.

2.2 ZrO

2

Sensor

2.2.1 Principles of Operation

Teledyne’s zirconium oxide sensor is a miniature solid state sensor.

It uses a stabilized zirconia disc as an electrolyte with a sensing
electrode (the cathode) and a counter electrode (the anode) on each side
of the disc. The zirconia is heated to a temperature of about 500°C via a
built-in heater. By applying a suitable reference voltage (approximately
0.75 volt) across the cathode and anode, the heated zirconia acts as an
electrolyte which is capable of moving oxygen ions within its crystalline
structure. This allows for the reduction of oxygen molecules reaching
the cathode and the generation of a current equivalent to the parts per
million oxygen in the sample gas.

This sensor has been modified to sense low levels of oxygen—less

than 1 ppm and generate a current proportional to the concentration. The
sensor is designed for long term stability and an operating life of 5 years.