4 system self diagnostic test – Teledyne 3000ZA - Trace oxygen analyzer User Manual

Page 66

background image

Maintenance

3000ZA

Teledyne Analytical Instruments

54

2. To change between American and European fuses, remove

the single retaining screw, flip Fuse Block over 180 degrees,
and replace screw.

3. Replace fuse as shown in Figure 5-3.
4. Reassemble Housing as shown in Figure 5-2.

American Fuses

European Fuses

Figure 5-3: Installing Fuses

5.4 System Self Diagnostic Test

1. Press the System button to enter the system mode.
2. Use the __ arrow keys to move to More, and press Enter.
3. Use the __ arrow keys to move to Self-Test, and press Enter.

The following failure codes apply:

Table 5-1: Self-Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed