Maintenance – Teledyne 3000ZA - Trace oxygen analyzer User Manual

Page 64

background image

Maintenance

3000ZA

Teledyne Analytical Instruments

52

Maintenance

5.1 Routine Maintenance

Aside from normal cleaning and checking for leaks at the gas

connections, routine maintenance is limited to replacing the sensor and
fuses, and recalibration. For recalibration, see Section 4.4 Calibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.2 Cell Replacement

The ZrO

2

sensor is a sealed transducer with no electrolyte to

change or electrodes to clean. When the cell reaches the end of its useful
life, it is replaced. The spent sensor should be discarded according to
local regulations.

5.2.1 Removing the Sensor

The ZrO

2

sensor is located inside the nylon cell block behind the

front panel.

To remove an existing cell (refer to Figure 5-1):

1. Remove power to the instrument by unplugging the power

cord at the power source.

2. Open the front panel door by pressing the release button on

the top right corner of the door all the way in with a narrow
gauge tool.

3. Unscrew the collar on the top assembly of the cell block from

the bottom section of the cell block assembly. The sensor is
attached to pin sockets on the bottom assembly.