Trace Oxygen Analyzer

Operation

Teledyne Analytical Instruments

43

Use the __ arrow keys to enter the oxygen-concentration mode.

Use the __ arrow keys to blink the digit you are going to modify. Use
the __ arrow keys again to change the value of the selected digit. When
you have finished typing in the concentration of the span gas you are
using, press Enter to begin the Span calibration.

#### ppm Span
Slope=#### ppm/s

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates
information on Slope. Spanning automatically ends when the span
output corresponds, within tolerance, to the value of the span gas
concentration. Then the instrument automatically returns to the analyze
mode.

4.4.2.2 M

ANUAL

M

ODE

S

PANNING

Press Span to start the Span function. The screen that appears

allows you to select whether the span calibration is to be performed
automatically or manually.

Span: Settling:MAN
For Next

Use the __ keys to toggle between

AUTO

and

MAN

span settling.

Stop when

MAN

appears, blinking, on the display. Press Enter to move

to the next screen.

Press Enter to move to the next screen.

Calib. Holding time
Cal hold: 5 min

This menu allows the operator to set the time the analyzer should

be held in the AUTO span mode. It does not have any effect in the
MANual mode. Just press Enter key to continue.

Span Val: 000008.00
Span Mod #

Press _ () to permit modification (

Mod #

) of span value.

Use the arrow keys to enter the oxygen concentration of the span

gas you are using. The __ arrows choose the digit, and the __ arrows
choose the value of the digit.