Trace Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

Power: Universal power supply 85-250 V ac, at

47-63 Hz.

Operating Temperature: 5-35 °C

Accuracy: ±2% of full scale at constant temperature

(at ca;obrated ramge).
±5% of full scale over operating
temperature range, except 0-1 ppm
analysis range, once thermal equilibrium
is reached.
±0.2 ppm on 0-1 ppm analysis range, once
thermal equilibrium is reached.

Analog outputs: 0-1 VDC percent-of-range,

0-1 VDC range ID.
4-20 mA DC (isolated) percent-of-range,
4-20 mA DC (isolated) range ID
(optional).

Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep

(6.96 in high, 8.7 in wide, 12.2 in deep).