Trace Oxygen Analyzer

Operation

Teledyne Analytical Instruments

41

The zeroing process will automatically conclude when the output is

within the acceptable range for a good zero. Then the analyzer
automatically returns to the Analyze mode.

4.4.1.2

M

ANUAL

M

ODE

Z

EROING

Press Zero to enter the Zero function. The screen that appears

allows you to select between automatic or manual zero calibration. Use
the __ keys to toggle between

AUTO

and

MAN

zero settling. Stop when

MAN

appears, blinking, on the display.

Zero: Settling: Man
To Begin

Press Enter to begin the zero calibration. After a few seconds the

first of five zeroing screens appears. The number in the upper left hand
corner is the first-stage zero offset. The microprocessor samples the
output at a predetermined rate. It calculates the differences between
successive samplings and displays the rate of change as Slope= a value
in parts per million per second (ppm/s).

#### ppm Zero
Slope=#### ppm/s

Note: It takes several seconds for the true Slope value to display.

Wait about 10 seconds. Then, wait until Slope is
sufficiently close to zero before pressing
Enter to finish
zeroing .

Generally, you have a good zero when Slope is less than 0.05

ppm/s for about 30 seconds. When Slope is close enough to zero, press
Enter. In a few seconds, the screen will update.

Once zero settling is completed, the information is stored in the

microprocessor, and the instrument automatically returns to the Analyze
mode.

4.4.1.3

C

ELL

F

AILURE

Cell failure in the 3000ZA is usually associated with inability to

zero the instrument down to a satisfactorily low ppm reading. When this
occurs, the instrument will still Zero, but the 3000ZA system alarm
trips, and the LCD displays a failure message.

#.# ppm Anlz