Teledyne 3000ZA - Trace oxygen analyzer User Manual
Page 34
Installation
3000ZA
Teledyne Analytical Instruments
22
Span: Floating input. 5 to 24 V input across the + and – pins puts the
analyzer into the Span mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Span mode to terminate when done. A synchronous signal
must open and close external span valve appropriately. See
Figure 3-5 Remote Probe Connector. (The –C option internal
valves operate automatically.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Table 3-3: Remote Calibration Connections
Pin
Function
9
+ Remote Zero
11
– Remote Zero
10
+ Remote Span
12
– Remote Span
40
Cal Contact
41
Cal Contact
Remote Calibration Protocol: To properly time the Digital Remote Cal
Inputs to the Model 3000ZA Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote
Cal Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that
request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.
For example:
1. Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2. When the CRC closes, remove the zero command.