Operation

3000ZA

Teledyne Analytical Instruments

48

L—10 M—100
H—250 Mode—AUTO

Use the __ arrow keys to blink the range to be set: low (L), medium

(M), or high (H).

Use the __ arrow keys to enter the upper value of the range (all

ranges begin at 0 ppm). Repeat for each range you want to set. Press
Enter to accept the values and return to Analyze mode. (See note
below.)

Note: The ranges must be increasing from low to high, for

example, if range 1 is set as 0–10 ppm and range 2 is set
as 0–100 ppm, range 3 cannot be set as 0– 50 ppm since
it is lower than range 2.

4.6.2

Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing

the analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range

function by pressing the Range button on the front panel.

Use the __ arrow keys to move the blinking over

AUTO

.

Use the __ arrow keys to switch from

AUTO

to

FX/LO

,

FX/MED

, or

FX/HI

to set the instrument on the desired fixed range (low, medium, or

high).

L—10 M—100
H—250 Mode—FX/LO

—or—

L—10 M—10 0
H—250 Mode—FX/MED

—or—

L—10 M—10 0
H—250 Mode—FX/HI

Press Escape to re-enter the Analyze mode using the fixed range.

Note: When performing analysis on a fixed range, if the oxygen

concentration rises above the upper limit (or default value)