Model 2230r, process hydrogen analyzer – Teledyne 2230R - Process Hydrogen Analyzer User Manual

Page 26

background image

MODEL 2230R, Process Hydrogen Analyzer

Page 26

21

Settle time: 30 min

(Y/N)?

Type “ y”

22

Apply 30.000%H2:

Ready (Y/N)?

Type “ y”

Streaming data…

23

Taking Average…
cap=0.00000

res=x.xxxxx

Calibration Gas #2 finished.
Calibration complete.

4) Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of

Calibration, conduct the Verification sequence as described in the previous section to

verify that the calibration process was executed correctly.