MODEL 2230R, Process Hydrogen Analyzer

Page 23

FIELD VERIFICATION STEPS

Step

Display

User response

1

H2scan:

Type “ e” to run field verification

2

Ready to Verify

(Y/N)?

Type “ y

3

Pressure is 1.000

atm (Y/N)?

Type “ n

4

Enter new pressure:

Type “ .969” for the local atmospheric pressure

5

Pressure is 0.9690

atm (Y/N)?

Type “ y

6

Save as default

(Y/N)?

Type “ n

7

Verify Sensor

(Y/N)?

Type “ y

8

Enter Gas (%H2):

Type “ 1” for gas #1, 1% hydrogen by volume

9

Settle time (min):

Type “ 30” for a 30 minute duration for gas #1

10

Tolerance (%H2):

Type “0.3%” absolute from the for 1% hydrogen by volume

11

Apply 1.000%H2:

Ready (Y/N)?

Type “ y

Streaming data…

12

Taking Average…

An average will be calculated. Verification Gas1 finished

13

Continue test Verify

Sensor (Y/N)?

Type “ y

14

Enter Gas (%H2):

Type “ 30” for gas #2, 30% hydrogen by volume

15

Settle time (min):

Type “ 30” for a 30 minute duration for gas #2

16

Tolerance (%H2):

Type “1.0%” absolute from the TOLERANCE CHART for 30%

hydrogen by volume

17

Apply 30.000%H2:

Ready (Y/N)?

Type “ y

Streaming data…

18

Taking Average…

An average will be calculated. Verification Gas2 finished.

19

Continue test Verify

Sensor (Y/N)?

Type “n” to end field verification.

Verification complete.

4).

If the unit passes Verification, Calibration is not required at this time.

5).

If the unit fails Verification, you will see the following:

18

Taking Average…

An average will be calculated.

19

Verify Failed

Verification complete.

6).

If the unit fails Verification, TELEDYNE ANALYTICAL INSTRUMENTS recommends

Calibration, detailed in the following section.