beautypg.com

Model 2230r, process hydrogen analyzer – Teledyne 2230R - Process Hydrogen Analyzer User Manual

Page 17

background image

MODEL 2230R, Process Hydrogen Analyzer

Page 17

6 – Temperature calibration data

A – All of the above

E

Field Verify (refer to the Section 5.4 VERFICATION).

F

Field Calibration (refer to the Section 5.5 CALIBRATION).

G < fmt>

Start or resume the sensor operation: if needed, heat the ASIC, setup

the sensor, and output data on serial port; this restores default

settings (refer to the FORMAT and OPTIONS sections).

H

Set the hydrogen reporting range: to in %H₂.

I

Set the DAC current output range: to ; error output
; and not ready output milliamps. Possible range from

0 to 20 milliamps.

L < fmt>

Print current hydrogen reading. Used to poll for hydrogen readings.

Default format is current setting (refer to the FORMAT

and OPTIONS sections).

P

Select atmospheric pressure of gas. Factory default pressure is 1

ATM.

S

Stop the sensor: turn off heater, set Bias to zero, set DAC outputs to

zero, and stop reporting data on the serial output.

V

Set the DAC voltage output range: to ; error output

; and not ready output in volts. Possible range from 0

to 5 volts.

X

Clear field calibration data (returns to last factory calibration data).

Format

The Format string is a two character hexadecimal representation

of an 8 bit value derived from the following table. The user determines

which data is needed and selects that bit value. Once all selections are

made the values are summed bitwise and then converted to a two

place hexadecimal value. To aid in the conversion, a 4 bit to
hexadecimal conversion table follows in EXAMPLE 3.

Format: Bit Value Identifiers

Serial output Format parameter: select the desired columns of data from this list, add

the bit value for each column bitwise, and convert into two hexadecimal characters using the

4 bit-to-Hexadecimal table in EXAMPLE 3.

Description

Bit Value

Include time stamp

1

0

0

0

0

0

0

0

Include raw ADC values

0

1

0

0

0

0

0

0

Include PCB temperature

0

0

1

0

0

0

0

0

Include sensor temperature

0

0

0

1

0

0

0

0

Include capacitor reading

0

0

0

0

1

0

0

0