beautypg.com

Model 2230r, process hydrogen analyzer – Teledyne 2230R - Process Hydrogen Analyzer User Manual

Page 24

background image

MODEL 2230R, Process Hydrogen Analyzer

Page 24

5.5 CALIBRATION

Calibration Interval

Calibration is used to correct any offset that exist between the sensor output and a known

hydrogen concentration. Calibrations do not cause any wear on the sensor and can be

accomplished as often as desired. It is recommended that Calibration be performed if a unit

fails Verification.

Calibration Procedure

IMPORTANT NOTE: During the Calibration process any previously completed Field Calibrations

are cancelled. As a result, during the routine the unit may display a hydrogen concentration

that is different from the applied gas concentration. This is normal. Once the procedure is

completed, the readings will be corrected to display the right concentrations for all subsequent

exposures.

Calibration can only be accomplished through interface with the unit via the serial port (refer to
the previous Serial Communication Commands section and the LEVEL 1 COMMANDS table in

Section 4.3). Analog outputs can be monitored through the user’s system.

Calibration Using Serial Interface (firmware version 0.47) – Follow the sequence below:

1) Press “Esc

2) The unit will return the command prompt “H2scan:

3) Type “f” then hit “Enter” and follow prompts to field calibrate the sensor with two gas

concentrations

These steps will implement the following specification values as an example; for values in bold,

please substitute appropriate values relative to your specific operational conditions:

Local atmospheric pressure: 0.969 ATM

Hydrogen range: 0% to 30% hydrogen, balance nitrogen

Field Verification Gas #1: 1% hydrogen, balance nitrogen (instead of gas #1 being 0%

hydrogen, H2scan recommends Field Verification gases always have a hydrogen

concentration)

Field Verification Gas #2: 30% hydrogen, balance nitrogen

Settle Time: 30 minutes (TELEDYNE ANALYTICAL INSTRUMENTS recommends this

duration for field calibration)

WARNING:

AS IN THIS EXAMPLE, FOR OPTIMIZATION, THE HYDROGEN

CONCENTRATIONS OF THE GASES IMPLEMENTED IN FIELD CALIBRATION
AND FIELD VERIFICATION MUST BE THE SAME.