If the unit gets exposed to extended periods of no H₂ (or some Oxygen), condition the

sensor as described in the Start-up section and follow-up with a Field Verification (and Field

Calibration, if needed).

Effect of pressure: The Model 2230R Analyzer is hydrogen specific and sensitive to only the
hydrogen partial pressure in the gas stream. Since changes in total gas pressure will affect

the hydrogen partial pressure, they will also affect the sensor readings. For instance, at one

atmosphere pressure, a 50% H₂/N₂ mixture will be reported as 50% from the unit. At 1.1

atm, the reading will increase to 55% and two atms will result in a reading of 100%. In fact,

the Model 2230R Analyzer are capable of measuring multiple atmospheres of hydrogen and

readings above 100% H₂ are interpreted as hydrogen pressures above one atmosphere. So,

for example, a reading of 150% H₂ means 1.5 times the hydrogen pressure of a 100% H₂

concentration at one atmosphere. At the factory, the units are calibrated at one

atmosphere pressure. Performing the Field Calibration at the operating pressure will display

the pressure corrected reading. For example, if the local atmospheric pressure is 0.97 atm,
doing a Field Calibration will correct for this.

Verification/Calibration gas bottle accuracy: The inaccuracy of the gas bottle concentration

will directly affect the measured accuracy by the units. During factory calibration, the units

are calibrated with high accuracy gases (as high as ± 0.02% accurate). It is strongly

recommended that the user perform calibration with similar high accuracy gases to

maintain the accuracy specified in the manual.

4. ELECTRICAL & COMMUNICATION

4.1 CONNECTIONS

Power/Analog Output - Connector 1

Supplied Cable – 4m (12 ft.) standard length (Other lengths available)

Wire Color

Description

Brown

+8 VDC to +13 VDC

White

VDC Return

Black

Positive Analog Output

Blue

Analog Output Return

WARNING:

IF THE UNIT IS INSTALLED IN A CLASSIFIED LOCATION THEN IT

IS THE RESPONSIBILITY OF THE USER AND INSTALLER TO MAKE
CONNECTIONS TO RELATED EQUIPMENT IN A MANNER CONSISTENT

WITH THE LOCATION CLASSIFICATION.