Model No. PS-2172

Specifications

9

Appendix B: Reference Solutions for
Conductivity Calibration

Reference

Solution

Approximation

Normality of

Solution

Method of Preparation

Temperature

(°C)

Electrical

Conductivity

1

(µS/cm)

A

1

74.2480 g of KCI weighed in air per 1 L of
solution at 20 °C

0

65176

18

97838

25

111342

B

0.1

7.4365 g of KCI weighed in air per 1 L of
solution at 20 °C

0

7138

18

11167

25

12856

C

0.01

0.7440 g of KCI weighed in air per 1 L of
solution at 20 °C

0

773.6

18

1220.5

25

1408.6

D

0.001

Dilute 100 mL of solution C to 1 L at 20 °C

0

77.69

18

127.54

25

148.93

1

Excluding the conductivity of the water used to prepare the solutions. These tabulated conductivity values are in international

units. When using measuring instruments calibrated in absolute units, multiply the tabular values by 0.999505