beautypg.com

Specifications – PASCO PS-2172 Advanced Chemistry Sensor User Manual

Page 7

background image

®

Model No. PS-2172

Specifications

7

Specifications

General

Component Sensors

Temperature, pH/ISE/ORP probe voltage, Absolute Pressure, Voltage

Max. Sampling Rate

100 samples per second (for each component sensor)

Default Sampling Rate

10 samples per second

Temperature

Range

35 °C to 135 °C

Accuracy

0.5 °C

Resolution

0.01 °C or better

Sensing Element

10 k

 thermistor located in probe tip

pH/ISE/ORP probe voltage

Electrode Connector

standard BNC

Voltage Range

2000 mV to 2000 mV

Voltage Resolution

0.1 mV

pH Range

0 to 14

pH Resolution

0.001

Absolute Pressure

Range

20 kPa to 400 kPa

Accuracy

2 kPa

Resolution

0.1 kPa

Repeatability

1 kPa

Conductivity

Range

three user-selectable ranges:
0 to 1000

S/cm, 0 to 10000 S/cm, 0 to 100000 S/cm with included 10x probe

0 to 100

S/cm, 0 to 1000 S/cm, 0 to 10000 S/cm with optional 1x probe

Accuracy

±10% of full scale without calibration

Resolution

better than 0.05% of full scale