The parameters available are:

“F1” “Unit”

Default Weighing Unit

“F2” “dPP”

Decimal Point Position

“F3” “CAP”

Select capacity and increment

“F4” “CAL”

To enter the Calibration

“F5” “init” “ZEro”

Initial Zero Range

“F6” “rE” “ Zero”

Re-Zero range

“F7” “SCSivE” “ tArE”

Successive Tare Enable

“F8” “Ad” “CoUnt”

Display the A/D counts

“F9” “Pin”

To set the password

“F10” “Lvd” “mode”

Low voltage detection

“F11” “trACk” “ZEro”

Zero track setting

12.2.1

F1 –DEFAULT WEIGHING UNIT

To set the weighing unit that will be used initially for calibration and setting the
calibration. This weighing unit will always be available to the user.

• To enter this parameter, press the [Tare] key when “F1” is shown. The

display will show the current setting.

• Press the [U.Wt./Units] key to change the value. Select from the units

displayed, typically kilograms (kg), grams (g) or pounds (lb) for USA
sales and kilogram or grams for the rest of the world.

• Press [Tare] to accept the displayed value.

12.2.2

F2 –DECIMAL POINT POSITION

• To set the value for the decimal point location. The options are 0, 0.0,

0.00, 0.000, 0.0000

• To enter this parameter, press the [Tare] key when “F2” is shown.

• The display will show the current setting.