0 error, Codes – Nova-Tech GC Indicator User Manual
Page 29
P a g e
27
© Adam Equipment Company 2010
11.0 ERROR
CODES
During the initial power-on testing or during operation, the scale may show an error
message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message, turning the
balance on, carry out the calibration or other functions. If the error message is still
shown contact your dealer for further support.
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 1
Time input error.
Tried to set an illegal time, i.e.
26hours
Err 2
Date input error
Tried to set an illegal date, i.e. 36th
day
Err 4
Initial Zero is greater
than allowed (typically
4% of the maximum
capacity) when power is
turned on or when the
[Zero] key is pressed,
Weight is on the pan when turning
the scale on.
Excessive weight on the pan when
zeroing the scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct
when turning the scale
on.
Platform is not installed.
Damaged Load cell.
Damaged Electronics.
diSPEr
Shows when a number
is too large tot fit the
display
The accumulated totals for weight
or count exceed 999999. The data
on RS232 is still shown accurately