beautypg.com

Nova-Tech GC Indicator User Manual

Page 31

background image

P a g e

29

© Adam Equipment Company 2010

12.1.1

F1 -CALIBRATION

To enter the calibration parameter, press the [Tare] key when “F1” “CAL” is displayed.
The scale should be calibrated using a mass of approximately 2/3Maximum to
Maximum. If the scale has been calibrated previously the value will be stored.
Details of calibration are given is section 8.

12.1.2

F2 –PIN

• To enter this parameter, press the [Tare] key when “F2” “Pin” is shown.

• The “Weight” display will show “Pin 1”

• Enter the new password. The “Unit Weight” display will show dashes.

Press [Tare].

• The “Weight” display will change to “Pin 2 “, Enter the password again

and press [Tare].

• The display will show “donE” indicating the new password has been

accepted and will return to the menu. “F3” “Unit” is shown.

• Record the new password in a secured place.

• Press [Zero] to return to weighing.

12.1.3

F3 –Unit

• To enter this parameter, press the [Tare] key when “F3” “unit” is

shown.

• This parameter allows the user to enable or disable the weighing units

available in his scales.

• The display will show the first available weighing unit and if it is On or

Off.

• Press the [U.Wt./Units] key to change the value. Press [Tare] to

accept the displayed setting and go to the next unit.

• Press the [Zero] key to return to the menus. Press the [Zero] key again

to return to normal weighing.