Stimulus Generator 4000 Series Manual
66
Please make sure to use the correct format.
1. Connect the digital IN /OUT connector to a TTL signal source.
2. Ground digital inputs manually which are not used.
3. Set up your own stimulus protocols, either nine stimulus patterns or up to 256 patterns,
and save the files.
4. On
the
Edit menu, click Trigger Settings.
5. Select the option Extended Multi-File.
6. Optionally select the check box "Show all Patterns".
7. Define the desired files to each of the positions, or load a configuration file (*.mf2)
for the “Extended Multi File” mode.
8. Close the dialog by clicking "OK".
9. All selected files are opened automatically.
10. Check all stimulus protocols before download. Please make sure that you have selected
all necessary channels under Download Channel for each file separately.
11. Download the files onto the STG by clicking Download on the STG menu.
12. The stimulation as specified in the selected MC_Stimulus II files will now be triggered