The analysis ROI covers the whole duration of the voltage (test) pulse except the first 10 ms in
order to avoid interference with the capacitive transient caused by the voltage jump.
Next, the recording is started.
The sequence generates the following voltage jump:
a. Start the recording
b. 100 ms at -60 mV
c. 200 ms at +50 mV
d. 200 ms at -60 mV
e. Stop the recording
Finally, the current amplitude elicited by the (in this example depolarizing) voltage pulse is sent to
the log window, and
the signal is compared to the user-defined limits minResponse and maxResponse. If the elicited
signal is positive (outward current), the condition
if (signal < minResponse || signal > maxResponse)
should be used.
If the expected signal is negative current (inward current) the operators > and < have to be
exchanged.
Voltage Protocols (IV-dependencies)
- MATLAB Manual (21 pages)
- MEA-IT Manual (30 pages)
- MEA-VMT Manual (17 pages)
- USB-ME64-128-256-Systems Manual (36 pages)
- LTP-Director Manual (89 pages)
- MPA2I Manual (12 pages)
- MPA32I Manual (39 pages)
- MPA8I Manual (22 pages)
- MEA2100-System Manual (72 pages)
- nanoZ manual (45 pages)
- Wireless-System Manual (62 pages)
- µPA16+32 Manual (21 pages)
- Portable_ME-System Manual (45 pages)
- MEA2100-System (6 pages)
- MEA2100-System (2 pages)
- STG4002 Manual (110 pages)
- MEA2100-CO2-C (1 page)
- USB-MEA256-System Manual (66 pages)
- MEA Manual (135 pages)
- PGA Manual (29 pages)
- TC01-2 Manual (30 pages)
- STG1000 Manual (89 pages)
- PH01 Manual (10 pages)
- ME-System Manual (31 pages)
- STG3008-FA Manual (101 pages)
- STG-Lite Manual (27 pages)
- USB-MEA32-STIM4 Manual (50 pages)
- HiClamp Manual (81 pages)
- MEAS21+MEASwitchUSB Manual (23 pages)
- MEAS42+MEASwitchUSB Manual (23 pages)
- MEA_Signal_Generator Manual (29 pages)
- MEA1060-Inv Manual (35 pages)
- MEA1060-Inv-BC Manual (64 pages)
- MEA1060-Up Manual (36 pages)
- MEA1060-Up-BC Manual (64 pages)
- Roboocyte2 JavaScript Manual (5 pages)
- PS40W Manual (9 pages)
- PPS2 Manual (26 pages)
- STG4004+STG4008 Manual (114 pages)
- Roboocyte2 Manual (115 pages)
- TC01-TC02 Manual (33 pages)
- Roboinject Manual (91 pages)
- MEA-System Manual (28 pages)
- MC_Rack Manual (201 pages)