Multichannel Systems Roboocyte2 Scripting Manual User Manual
Page 24

Roboocyte2 Manual
24
If the electrode resistances are within the range (between MIN_RESISTANCE and
MAX_RESISTANCE) the script proceeds; if not (else), the script breaks and continues with the
next oocyte.
Now, after completing electrode offset compensation, oocyte impalement can be started.
2.3.5 Oocyte Impalement
The oocyte impalement is performed by using the command:
Robo2.Impale(MIN_RMP, IMPALEMENT_STEPS, IMPALEMENT_STEP, IMPALEMENT_WAIT)
MIN_RMP, IMPALEMENT_STEPS, IMPALEMENT_STEP and IMPALEMENT_WAIT are predefined
variables (see table in chapter 1.2.3).
The impalement process is started by a movement of the z-axis to the default oocyte impalement depth
(default = 800 µm above the well plate bottom). Then n =
IMPALEMENT_STEPS are performed with a
step size of
∆
z = IMPALEMENT_STEP and a waiting time t = IMPALEMENT_WAIT between
individual steps.
Default values for IMPALEMENT_STEPS and IMPALEMENT_STEP are 8 and 50 µm, respectively.
- MATLAB Manual (21 pages)
- MEA-IT Manual (30 pages)
- MEA-VMT Manual (17 pages)
- USB-ME64-128-256-Systems Manual (36 pages)
- LTP-Director Manual (89 pages)
- MPA2I Manual (12 pages)
- MPA32I Manual (39 pages)
- MPA8I Manual (22 pages)
- MEA2100-System Manual (72 pages)
- nanoZ manual (45 pages)
- Wireless-System Manual (62 pages)
- µPA16+32 Manual (21 pages)
- Portable_ME-System Manual (45 pages)
- MEA2100-System (6 pages)
- MEA2100-System (2 pages)
- STG4002 Manual (110 pages)
- MEA2100-CO2-C (1 page)
- USB-MEA256-System Manual (66 pages)
- MEA Manual (135 pages)
- PGA Manual (29 pages)
- TC01-2 Manual (30 pages)
- STG1000 Manual (89 pages)
- PH01 Manual (10 pages)
- ME-System Manual (31 pages)
- STG3008-FA Manual (101 pages)
- STG-Lite Manual (27 pages)
- USB-MEA32-STIM4 Manual (50 pages)
- HiClamp Manual (81 pages)
- MEAS21+MEASwitchUSB Manual (23 pages)
- MEAS42+MEASwitchUSB Manual (23 pages)
- MEA_Signal_Generator Manual (29 pages)
- MEA1060-Inv Manual (35 pages)
- MEA1060-Inv-BC Manual (64 pages)
- MEA1060-Up Manual (36 pages)
- MEA1060-Up-BC Manual (64 pages)
- Roboocyte2 JavaScript Manual (5 pages)
- PS40W Manual (9 pages)
- PPS2 Manual (26 pages)
- STG4004+STG4008 Manual (114 pages)
- Roboocyte2 Manual (115 pages)
- TC01-TC02 Manual (33 pages)
- Roboinject Manual (91 pages)
- MEA-System Manual (28 pages)
- MC_Rack Manual (201 pages)