beautypg.com

Multichannel Systems STG4002 Manual User Manual

Page 89

background image

Analog Output Signals

85

Example 1b: Current pulse with programmed 1600 μA for 20 μs and 20 μs interval and 10 k

load resistance.

The pulse is biphasic, and 1600 μA which is the limit of the maximum capacity of the STG 4002.
Green trace = current output (+I), pink trace = TTL trigger output from the Sync Out.

Result of Example 1: The current output signal is delivered unmodified without any influence
of the length of the pulse with an amplitude of 1600 μA which is at the maximum capacity of
the STG 4002.