beautypg.com

Multichannel Systems STG4002 Manual User Manual

Page 24

background image

Stimulus Generator 4002 Manual

20

In the following example, the stimuli on output channels 1 and 2 triggered by Trigger 1 are
repeated three times after the trigger event, each time accompanied by the Sync Out 1 output.
All other stimuli are applied once after their respective trigger event.

Repeating complete pulse protocols, Gating Trigger

Repeated stimulation of complete protocols can also be controlled by a gating trigger. Channels
associated with a Trigger input will be active as long as the condition of that Trigger is 1 (HIGH)
and switched off as soon as the condition becomes 0 (LOW). Gating trigger function can be
enabled in the Trigger Settings menu.