beautypg.com

MP Pumps CHEMFLO 5, 6, 7, 8 PUMPAKS INSTRUCTION User Manual

Page 2

background image

2

ROTATION: (cont.)

To adjust impeller clearance, do this:

1.

Loosen the impeller drive sleeve clamp.

2. Move impeller either forward or back by

using a screwdriver to push impeller

back – or move impeller drive sleeve

forward.

Depending on the seal spring tension, you can do

this by hand without removing the pump housing.

Turn the driver to determine that you have

adjusted the clearance so the impeller does not

rub. Then tighten the impeller drive sleeve

clamp.

IF THE SEAL SPRING TENSION

PREVENTS

THIS

METHOD

OF

ADJUSTMENT, DO THIS:

1.

Remove pump housing.

2.

Loosen drive clamp, but do not remove.

3.

Remove o-ring from housing.

4.

Replace housing, pushing against impeller

face. Secure housing with two fasteners, 180

degrees apart.

5.

Securely tighten drive clamp.

6.

Remove housing and install o-ring

7.

Replace housing.

INSTALLATION:

The MP CHEMFLO centrifugal PUMPAKS

should be installed with flooded suction or as

near to the liquid source as possible.

If it is necessary to install the pump above the

level of the liquid, a priming line should be

connected directly to the suction line. A check

valve must be used in conjunction with a priming

line or the priming liquid will merely pass

through the suction line to the liquid source. The

priming of the suction line evacuates the air in

the suction line and pump. This series of pumps

will not handle air and therefore all air must be

displaced from the suction line and pump before

the pump can operate. Be sure that the pump

housing is filled with liquid before starting the

pump.

THE MECHANICAL SEAL IN THE

PUMP MUST NOT BE OPERATED DRY.

Pipe or hose of the same size (or larger) as inlet

and outlet openings should be used on the

installation. When using pipe, avoid sharp bends

and use long radius elbows wherever possible.

This will keep friction loss at a minimum and

allow the pump to operate more effectively. Use

pipe dope on all connections and be sure all

fittings are airtight, especially on the suction side

of the pump. An air leak on the suction side of

the pump will prevent proper operation. A

section of non-collapsible hose between piping

and pump may be used as a vibration dampener.

TO DISASSEMBLE THE PUMP:

Drain the system of liquid, break suction and

discharge pipe unions, and, if necessary, remove

all piping from the suction and discharge

openings. Remove the fasteners holding the

pump adaptor to the driver, loosen the drive

clamp assembly, and remove the

PUMPAK.

To disassemble, remove the washers and hexnuts

holding the motor adaptor to the housing.

Remove the housing.

Loosen the drive clamp assembly. The impeller,

drive sleeve, seal bellows, and the spring

assembly will now slide forward free of the pump

adaptor.

2