Meriam M203 Altimeter Indicated Air Speed Tester User Manual
Page 18
15
Lockout Select
Enabling the Lockout feature prevents unauthorized users from making changes to
the configuration of the manometer. To enter the Program Mode, the user must
first enter the “password” (two-digit Lockout Code) within approximately 40
seconds when prompted. Failure to enter the correct two digit code within
approximately 40 seconds will return the unit to Measure Mode. Any two-digit
numeric code can be programmed. The factory Lockout Code of 00 (which is the
default as shipped from the factory) disables the Lockout.
To set the Lockout Code follow these steps:
Keystroke
Display
1. From the Measure Mode
press the PRGM key. If the
Lockout is set, enter the
correct “password” when
prompted.
Top line reads “PROGRAM
MODE” and bottom line reads
“UNITS SELECT”.
2. Press the up arrow key
twice.
Bottom line reads “USER
INFO SELECT”.
3. Press the right arrow key
then the up arrow key four
times.
Top line reads “LOCKOUT
CODE” and bottom line reads
“ENTER TO SELECT”.
4. Press the right arrow key,
then press the up arrow keys
to change the first digit. Press
the right arrow key to proceed.
Bottom line shows the old
Lockout Code. The cursor
flashes at the first position
while the value is changed,
the cursor moves to the right
position once the right arrow
key is pressed.
5. Press the right arrow key
when the desired code is set.
Lockout is activated.
Top line reads “LOCKOUT
CODE” and bottom line reads
“ENTER TO SELECT”.
6. Press the left arrow key
twice
Returns to Measure Mode.