CALIBRATION FOR RATE

With each flow transmitter there will be a calibration sheet that will supply the average pulses per volume
of specific units. Using the formulas below calculate the KFR value to enter into the display.

KFR VALUE CALCULATIONS FOR RATE

Gallons per Minute:

6000 / Average K-Factor in Pulses per Gallon

Liters per Minute:

6000 / (Average K-Factor in Pulses per cc X 1000)

Pounds per Minute:

6000 / (Average K-Factor in Pulses per Gallon / (Specific Gravity X 8.35)

Kilograms per Minute:

6000 / (Average K-Factor in Pulses per cc X 1000) / Specific Gravity

ENTERING THE KFR FACTOR FOR RATES

1. Cycle display using the MODE button, until the

RATE A

is displayed.

2. Press and hold the ENT button until the display enters program mode. At this point you should

see

KFR A VALUE

, with a number underneath it. The first digit should be blinking.

3. Using the UP or DN and SEL buttons adjust the

KFR A VALUE

to figures calculated from

above.

4. Press ENT

to accept

KFR A VALUE

setting. This will then move to the next screen.

5. Press ENT

to accept

SAMPLE A

setting.

6. Note: Do not adjust the SAMPLE A SETTING!

7. Press ENT

to accept

ENG UNITS A

setting.

8. Note: Do not adjust the SAMPLE A SETTING!

9. Calibration for

RATE A

is now complete.