beautypg.com

Calibration for ratios – Magnum Venus Plastech UPMS-1000 Monitoring System User Manual

Page 23

background image


OPERATIONS MANUAL

UPMS-1000-XXX

MAGNUM VENUS PLASTECH

REV 1.02

NOVEMBER 2010

PAGE 23

CALIBRATION FOR RATIOS


1. With the ratio block assembly on the front of the dispensing gun, turn on the gun, wait for pressures to

balance and come up to operating pressures.

2. Stop the pump system.

3. Cycle the UPMS-1000 display using the MODE

button until

TOTAL A

is displayed.

4. Press the RESET

button to zero the display.

5. Press the MODE

button to display

TOTAL B

6. Press the RESET

button to zero the display.

7. Start the pump system

8. Dispense both A and B materials into pre weighed cups.

9. Stop the pump system.

10. Find the net weight of both materials.

11. Cycle the UPMS-1000 display using the MODE

button until

TOTAL A

is displayed.

12. Press and hold the ENT button until the display enters program mode. At this point you will see

KFT

A VALUE

, with a number underneath it. The first digit should be blinking.

13. Using the UP or DN and SEL buttons adjust the

KFT A VALUE.

14. Press the ENT button twice to return to

TOTAL A

display. Verify the displayed value of the

TOTAL A

matches what was weighed in the cup. If the display does not match, return to step 9 and

repeat the process until the displayed value matches the actual value.

15. Once the TOTAL A display has been calibrated, press the ENT button to cycle the display to

TOTAL B

16. Press and hold the ENT button until the display enters program mode. At this point you will see

KFT

B VALUE

, with a number underneath it. The first digit should be blinking.

17. Using the UP or DN and SEL buttons adjust the

KFT B VALUE.

18. Press the ENT button twice to return to

TOTAL B

display. Verify the displayed value of the

TOTAL B

matches what was weighed in the cup. If the display does not match, return to step 13

and repeat the process until the displayed value matches the actual value.