1 program steps – Liquid Controls IT375 User Manual
Page 12
Sponsler, Inc.
Model IT375
Programming Pg. 12
4.1 PROGRAM STEPS
Step
Comment
CAL 00
Pulse Output (applies to DC Power/Alarm version only).
0 = No pulse output, low and high alarms.
1 = Scaled pulse output and low alarm.
2 = Unscaled pulse output and low alarm.
CAL 01
Kfactor - whole numbers.
CAL 02
Kfactor - digits after the decimal point.
The Kfactor is the pulses per unit of measure (eg. pulses/litre,
pulses/gallon, etc). The Kfactor can be programmed in the
range of 0.000001 - 999,999.
See Section 3.4.
CAL 03
Cutoff Frequency.
This determines the cutoff frequency in the range of 0.01 -
0.99Hz. The default setting is 0.25HZ.
Note that care must be taken when programming this value
because a low value may cause a slow update time.
CAL 04
Decimal Point for Rate Display.
The flow rate can be displayed with 0, 1, 2 or 3 places after the
decimal point.
CAL 05
Timebase for Rate.
The rate can be displayed in units per second, minute, hour or
day.
0 = second
1 = minute
2 = hour
3 = day
CAL 06
Filter.
The filter constant for filtering the input signal.
1 No filtering.
to
99 Very heavy filtering.
CAL 07
Decimal Point for Total Display.
The totals can be displayed with 0, 1, 2 or 3 places after the
decimal point.
CAL 08
Total Conversion Factor - whole numbers.
CAL 09
Total Conversion Factor - digits after the decimal point.
CAL 10
Low Alarm - whole numbers.
CAL 11
Low Alarm - digits after the decimal point.