Figure 6: 3a sanitary turbine flowmeter, Pg.11 – Liquid Controls Precision Turbine Flowmeters User Manual

Page 13

background image

FIGURE 6: 3A SANITARY TURBINE FLOWMETER

DISASSEMBLY FOR 3A SANITARY METERS

REASSEMBLY FOR 3A SANITARY METERS

1. Remove O-ring.

2. Push entire assembly out through the O-ring side of the meter.

3. Review all parts for damage

• Check bearings thoroughly

• Check O-rings for cracks and nicks

NOTE: If any part appears damaged, DO NOT REASSEMBLE. Call the factory for instructions.

NOTE: Internals must go back together the exact way that they come out.

1. Assemble the internal assembly as illustrated.

2. Push the assembly into the housing on the O-ring side.

3. Butt clip hanger up against the step inside the housing

4. Reinstall the O-ring.

5. Make sure the rotor spins freely and the clip hangers do not turn at all.

ITEM DESCRIPTION

QTY

1

HOUSING

1

2

ROTOR

1

3

CONE

2

4

HANGER

2

5

O-RING

1

6

SHAFT (Carbide or Teflon)

1

Pg.11