run Mode violation
When.the.CIM100.cover.is.removed.and.the.Chassis.Intrusion.button.releases,.the.CIM100.will.enter.an.

error.state..The.CIM100.display.will.flash.between.a.“violated”.message,.shown.on.the.left,.and.the.run.

mode.totalizer..Press.the.Select.button.to.proceed.to.the.Reset.Chassis.Intrusion.option.
reset chasiss Intrusion
The.Reset.Chassis.Intrusion.option.returns.the.CIM100.to.run.mode...

To RETuRn ThE CIM100 To Run MoDE:

1..Press.the.Menu.button.until.the.display.scrolls.to.the.Run.Mode.Violation.option..
2..Press.the.Select.button..The.display.will.show.the.close.message.and.begin.a.30.second.countdown..
3..Before.the.timer.expires,.reattach.the.CIM100.cover,.which.will.push.down.and.hold.the.Chassis.Intrusion.

button..When.the.timer.expires,.the.CIM100.will.enter.run.mode

reset totalizer
This.option.resets.the.resettable.totalizer.

without.using.an.external.magnet.

To REsET ThE REsETTAblE ToTAlIzER:

1..Press.the.Menu.button.until.the.display.scrolls.to.the.Reset.Totalizer.option.
2..Press.the.Select.button.to.toggle.between.“yes”.and.“no”..

c-factor
This.option.is.used.to.calibrate.the.CIM100..See the Calibration section on page 14.
totalizer decimal
This.option.sets.the.decimal.place.of.the.CIM100.totalizer..This.setting.has.an.indicator.light.on.the.run.

mode.display.

To sET ThE DECIMAl PlACE:

1..Press.the.Menu.button.until.the.display.scrolls.to.the.Totalizer.Decimal.option.
2..Press.the.Digit.button.to.change.the.decimal.place..Settings.include.0.00,.0.0,.0.(no.decimal),.X10,.X100.

totalizer units
This.option.sets.the.unit.of.measurement.for.the.CIM100.totalizers..This.setting.has.an.indicator.light.on.

the.run.mode.display.

To sET ThE unITs oF MEAsuREMEnT

:

1..Press.the.Menu.button.until.the.display.scrolls.to.the.Totalizer.Units.option.
2..Press.the.Digit.button.to.change.the.units.of.measurement..Settings.include.GAL=gallons,.LIT=liters,.

M3=cubic.meters,..=raw.pulse.count.data

.

Input frequency lower limit
This.option.sets.a.low.flow.cut.off.frequency..When.the.frequency.of.pulses.(in.Hz).drops.below.the.cut.

off.value,.the.CIM100.will.stop.registering.the.product.flow..See.the.test.sheet,.provided.by.the.factory,.for.

lower.limit.frequencies.

To sET low Flow CuT oFF vAluE:

1..Press.the.Menu.button.until.the.display.scrolls.to.the.Input.Frequency.Lower.Limit.option.
2..Press.the.Select.button.to.begin.editing.the.value..
3..Press.the.Digit.button.to.change.the.flashing.number..Press.the.Select.button.to.move.to.the.next.number..

Input frequency upper limit
This.option.sets.a.high.flow.cut.off.frequency..When.the.frequency.of.pulses.(in.Hz).surpasses.the.cut.off.

value,.the.CIM100.will.stop.registering.the.product.flow..See.the.test.sheet,.provided.by.the.factory,.for.

upper.limit.frequencies.

To sET hIgh Flow CuT oFF vAluE:

1..Press.the.Menu.button.until.the.display.scrolls.to.the.Input.Frequency.Upper.Limit.option.
2..Press.the.Select.button.to.begin.editing.the.value..
3..Press.the.Digit.button.to.change.the.flashing.number..Press.the.Select.button.to.move.to.the.next.number..

operatIon

Menu Mode options

12