Hach-Lange BIOGAS Basic User Manual User Manual
Page 50

English 50
Figure 6 Reagents menu
1
Title bar: indicates the name of the window.
5
GLP-Archives: access GLP tables and visualise 
the last reagent calibration results.
2
Calibrate/Enter titre: determine the concentra-
tion of the titrant (titre) by running a calibration or 
a calibration sequence or by entering the titre 
manually. 
6
Reagent library- supervisor use only: create, 
edit, reset and delete reagents to correspond to 
your specific needs.
3
Install reagents: install or replace reagents in a 
method or a sequence.
7
Reagent icon: Reagent status icon indicates the 
state of the reagent system. Four types of icons 
can be displayed. See table on the next page. 
4
Burette functions: fill, empty, flush, rinse and 
replace burette.
Table 1 Reagent icons
Icon
Description
Sunny icon: 
The reagent calibration or manual entry of the titre has been performed for all the 
reagents present in the system. Everything is just right!
Cloudy icon:
The reagent calibration of one of the reagents in the system should be performed within 
12 or 24 hours.
Stormy icon:
The reagent calibration of one of the reagents in the system has elapsed. 
At least one of the reagents present in the system has not been installed.
Question mark:
There is a problem in the editing of the reagent system. You need to be in Supervisor 
mode to solve the problem. Check the sequence/method parameters of the reagent or 
the electrode. 
Check the sequence or method, (press 1 in the Main window). The instrument indicates 
the possible errors and prompts you to correct them until ? disappears.
1
3
2
4
5
6
7