English, Specifications – Hach-Lange DR 3900 Basic User Manual User Manual
Page 103

English 103
English
Specifications
These are subject to change without notice!
Performance specifications
Operating mode
Transmittance (%), Absorbance and Concentration
Source lamp
Halogen lamp
Wavelength range
320–1100 nm
Wavelength Accuracy
± 1.5 nm (wavelength range 340–900 nm)
Wavelength
reproducibility
± 0.1 nm
Wavelength resolution
1 nm
Wavelength calibration Automatic
Wavelength selection
Automatic, based on method selection
Scanning speed
≥ 8 nm/s (in steps of 1 nm)
Spectral bandwidth
5 nm
Photometric measuring
range
± 3.0 Abs (wavelength range 340–900 nm)
Photometric accuracy
5 m Abs at 0.0–0.5 Abs
1 % at 0.50–2.0 Abs
Photometric linearity
< 0.5 % to 2 Abs
≤1 % at > 2 Abs with neutral glass at 546 nm
Stray light
< 0.1 % T at 340 nm with NaNO
2
Data log
2000 readings (Result, Date, Time, Sample ID, User
ID)
User programs
100
Physical and environmental specifications
Width
350 mm (13.78 in)
Height
151 mm (5.94 in)
Depth
255 mm (10.04 in)
Mass
4200 g (9.26 lb)
Ambient operating
requirements
10–40 °C (50–104 °F), maximum 80 % relative
humidity (without condensate formation)
Ambient storage
requirements
–40–60 °C (–40–140 °F), maximum 80 % relative
humidity (without condensate formation)
Additional technical data
Power connector via
external power supply
Input:100–240 V/50–60 Hz
Output: 15 V/40 VA
Interfaces
Use only shielded cable with maximum length of 3 m:
2× USB type A
1× USB type B
Use only shielded cable (e.g. STP, FTP, S/FTP) with
maximum length of 20 m:
1× Ethernet
Enclosure rating
IP40 (excluding interfaces and power supply)
Protection class
Class I
Performance specifications