Installation
3.5
Interfaces
The DR 3900 features three USB ports and one Ethernet port as
standard, located on the front and rear sides of the instrument
(
Figure 1
and
Figure 2, page 17
).
The USB type A ports are used for communications with a printer,
USB memory stick or keyboard. A USB memory stick can be
used to update the instrument software.
The USB type B port is used for communications with a PC. The
optional Hach Data Trans software (refer to
)
must be installed on the PC for data transfer.
A USB hub may be used to connect several accessories at a
time.
Note: USB cables must not be longer than 3 m.
These USB ports permit data to be exported to a printer or PC
and also allow the instrument software to be upgraded (refer to
). The Ethernet port supports real-time
data transfer in local networks, LIMS systems or SC controllers.
Only use a shielded cable (e.g. STP, FTP, S/FTP) with a
maximum length of 20 m for the Ethernet port.
3.6
Cell compartments, cell adapters and light shield
3.6.1
Cell compartments and adapters
The DR 3900 has two cell compartments (
Figure 3
). Only one cell
type at a time can be used for a reading.
Cell compartment (1) for:
•
13 mm round cells
Note: Cell compartment (1) contains barcode detection for
cells.
Cell compartment (2) for:
The following cell types can be used in cell compartment (2).
•
Without cell adapter A, 50-mm cells, 1-inch rectangular cells
and 1-inch flow cells can be used directly in cell compartment
(2).
•
With cell adapter A: 10 mm rectangular cells, 1-inch round
cells and AccuVac® vials.
Note: These cells must be inserted using cell adapter A.