Specifications
Specifications are subject to change without notice.
Analyzer
Specification
Details
Dimensions
Height: 450 mm; Width: 250 mm; Depth: 460 mm
Weight
7 kg (15.4 lb)
Casing protection
IP 65 / NEMA4X
Power supply
Standard version: 100-240 VAC 50/60 Hz
Low voltage version: 13-30 VAC 50/60 Hz, 18-42 VDC
Consumption: 25 VA
Measurement category: I (overvoltage less than 1,500 V)
Sample flow rate
20 liters/hour minimum
Sample tubing
Sample inlet and outlet: Diameter 8 mm (or 5/16'') semi-rigid tubing. We recommend the
use of PE tubing if sample temperature is inferior to 70 °C, and PTFE if superior to 70 °C
Connections
Power supply: Use the connector provided in the drawer
Analog output: Use the recommended POLYMETRON cable
Ambient temperature
-20 to 60 °C (-4 to 140 °F)
Maximum temperature 100 °C (at atmospheric pressure)
Maximum pressure
10 bar at ambient temperature
Relative humidity
10—90%
Precision
Conductivity: ± 2% of the displayed value
Temperature: ± 0.2 °C
Measurement range
Conductivity: 0.01 μS/cm to 200 μS/cm
Resistivity: 100 MΩ.cm to 5 kΩ.cm
Temperature: -20 to 200 °C (-4 to 392 °F)
Display resolution
0.001 μS/cm or 0.1 MΩ.cm