beautypg.com

Hach-Lange SPECTRO COLOR d-8 User Manual

Page 11

background image

BDA 387/2

12

B3.4.3 Gloss

subtraction

Select in the configuration menu

Settings

and

press key ok . Press key

Ø several times

to select

Gloss subtr. OFF

. Press key ok to

change the

Gloss subtr. OFF

. Now message

Gloss subtr. ON

appears and the gloss value is

subtracted in the calculation of the measuring

value.

To check the current settings, the top line in

colour and colour difference measurement modes reads D65/10/SPIN when the gloss

subtraction function is off (SPIN = specular gloss included) or D65/10/SPEX when the

gloss subtraction function is on (SPEX = specular gloss excluded).

B3.4.4 Turning beeper on and off

You can choose to activate or deactivate the

confirmation beep of spectro-color

®

.

Go to menu item

Beeper on

with the

Ø key

and confirm with ok .

B3.4.5 Pass / Fail Tolerance Setting

If you wish to employ the pass/fail function, you should first set the limit values for

ΔE*,

ΔL*,Δa*,Δb*. To do so, select item

Settings

in the configuration menu first and item

Pass/Fail Settings

next.

The following menu comes up:

The

ΔE*-limit is shown in inverse display. Set

the desired

ΔE*-limit using the × or Ø keys

and confirm with ok . The cursor jumps to the

first place after decimal point. The settings of

all other limit values are analogue.

Confirm the settings with ok .

Settings 

Ч 

Ш 

Date 
Time 
Language 

Gloss subtr. OFFFF 

Beeper on 
Pass/Fail Settings 
Back

 

ok 

Settings 

Ч 

Ш 

Date 
Time 
Language 
Gloss subtraction 

Beeper on 

Pass/Fail Settings 
Back

 

ok 

Pass/Fail Settings 

Ч 

Ш 

 ΔE* =  

1

0.00 

+ΔL* =  10.00 
ΔL* = ‐10.00 
+
Δa* =   1.00 
Δa* =  ‐1.00 
+
Δb* =   1.00 
Δb* =  ‐1.00

 

ok