beautypg.com

Maintenance - zero & purge – GxT V016-01, Ferret 16 GasLink II 5-Gas Analyzer User Manual

Page 13

background image

13

Maintenance - Zero & Purge

Zero

Use the SELECT buttons to move the selection arrow to select one

of the Calibration and Maintenance program.

When you select Zero, the analyzer will stop sampling gas from the

sample hose and will perform a zero of the bench. There is no need

to remove the sample hose from the tailpipe. Zeroing the analyzer

periodically will maintain the accuracy of the unit during testing.

Purge

Use the SELECT buttons to move the selection arrow to select one

of the Calibration and Maintenance program.

There are two Purge choices. One Purge function will purge for 1

minute, the other will purge for 5 minutes. Regardless of the purge

time selected, the analyzer will stop sampling gas from the sample

hose and will perform a purge of the bench and hose. There is no

need to remove the sample hose from the tailpipe. Purging clears

water and contaminants from the sample hose.

This manual is related to the following products: