Flowline LI40-1001 PodView User Manual
Page 12

12 of 16
MN204260
Rev A
GETTING AROUND PODVIEW®
Step Eight
PodView®  features  a  6‐digit  display  with  relay  indicators  and  a  three‐button  user  interface.    The  indicator 
displays the measured value and relay status of the connected EchoPod® (if the sensor has relays and they are 
configured for use). 
 Relay 1‐4 – Indicates when the sensor’s relay is
energized.  Note: Not all EchoPod® sensor models 
have relays. 
o
 If  the  EchoPod®  does  not  have  any  relays, 
then the indicators will remain off. 
 6‐digit Display – Shows the liquid level in height
(inches,  cm,  feet  or  meters)  or  the  volume  of 
liquid (gallons or liters). 
o
The selection of height vs. volume is set in the 
WebCal® software. 
 Up Button – Used to increase a set point value. 
 Select  Button  –  Used  to  enter  the  Menu  and 
accept values.
 Down Button – Used to decrease a set point
value.
Note:    To  increase  the  scrolling  speed  of  the  display, 
hold  down  the  SELECT  button  while  simultaneously 
pressing the UP or DOWN button. 
Entering the MENU:  If desired, users can change EchoPod® relay ALARM, VALVE or PUMP ON‐OFF set points 
using  PodView®.    To  enter  the  menu  PodView®  MENU  functions,  press  and  hold  the  SELECT  button  for  5 
seconds.  The MENU will then scroll between the configured PUMP, ALAMRS and RUN modes.  Note: If the 
relays are configured for PUMPS and ALARMS, then both will appear in the menu.  If the relays are configured 
for Alarms only, then PUMP will not appear.  If the relays are configured for PUMPS only (also for valves), then 
ALARMS will not appear. 
 To change a pump set point value, press SELECT when PUMP appears.   
 To change an Alarm set point value, press SELECT when ALARMS appear.   
 To exit the menu and return to run mode, press SELECT when RUN appears. 
