Flowline DL34 EchoPod User Manual

Page 22

background image

22 of 29

EchoPod

31 MAR 11

MN204210

Rev B

Tank Adapter

w/ 2”x1” Reducer Bushing

Tank Adapter

Thread x Thread

Do not use thread x thread

2. Riser: Installations with tall, narrow risers can impede the acoustic signal.

a. DL34 Series: 2” (5 cm) diameter risers should be no taller than 5” (12.7 cm).

Larger diameter risers should be no taller than 12” (30.5 cm).

b. DL14 & DL24 Series:

Note: Do not exceed the dimensions listed above

3. Flange (DL14 & DL24 series):

If installing on a flange, select a flange with a thread that is above the plane of the flange,
such as the LM52-1850.

a. The DL34 series works well with Flange installations.
b. Avoid the use of blind flanges with tapped threads or flanges where the threads are

even with the plane of the flange, such as the Banjo 1" Poly ANSI Flange (series
AF100).

c. Use a flange with a 2” thread and add a 2” to 1” reducer bushing to complete the

installation.

1” Flange w/

thread in plane

2” Flange w/

reducer bushing

Do not use thread in plane