Rev. 2/00 -4.5- 300/301 
the Fluid Sound Speed Compensation chart located 
in the Appendix of this manual. 
 
Once a new calibration frequency has been 
determined, the frequency is input as a 1 Vpp sine-
wave into TP3 on the main circuit card. This input will 
simulate full scale flow in the circuit. With the function 
generator inputting the calibration frequency the 
SPAN, 1mA, 20mA and CTR inputs can be 
configured. A 10% and 50% of SPAN frequency can 
be input and flow meter linearity can be verified. 
 
Noise and Gain Adjust 
 
If meter indicates a reading with no flow, turn R10 
(GAIN) C.C.W. just enough to eliminate. Since this 
reduces sensitivity, do not turn past point of proper 
operation. If indicator will not read flow, turn R10 
(GAIN) C.W. 
 
Relay Adjustment 
 
1. Turn R80 (DLY) fully CCW for minimum delay (2 
sec.) before proceeding with alarm trip 
adjustments. 
2. For a point of reference, turn R70 (ALM) also fully
CCW.
3. At the desired flow dropout (which can be
simulated by the procedure covered in the 
Electronic Calibration Procedure on Page 4.4), 
slowly (due to time delay) adjust alarm R70 (ALM) 
CW until relay is deactivated. Do not overturn. 
NOTE: About 10% flow hysteresis is built-in into 
the flow switch system at 1 FPS. 
4. After a desired alarm trip point is established,
adjust R80 (DLY) CW for a desired time delay (20 
sec. Max. On flow dropouts). 