Part 2 - transducer installation – Dynasonics D300_301 Series Doppler Ultrasonic Flow Meter User Manual
Page 12

 
Rev. 2/00 -2.4- 300/301 
permanently mounted. Under flowing liquid
conditions, adequate signal is indicated when the 
Signal LED (D21) flashes steadily. 
For proper operation, there cannot be air voids 
between the traducer face and pipe. The space must 
be filled with a material which is a good transmitter of 
sound energy such as: 
 
SILICONE GREASE: Dow Corning 111 R or 
comparable (-100 to +450 F.) The material must be 
suitable not to flow at temperature of pipe. Used for 
temporary survey installations and portable flow 
meters. 
 
SILICONE RUBBER: Dow Corning 732-RTV R. 
Excellent for permanent bonding. This adhesive is a 
recommended bonding agent and easily removable. 
 
 
INSTALLATION AND PIPE PREPARATION 
 
The cable from the DT6 transducer is provided with 
either dual-coaxial cables, flexible nylon conduit or 
PVC coated steel conduit with a 1/2” NPT fitting. The 
coaxial cable was ordered from the factory at a 
specific length UNDER NO CIRCUMSTANCES 
should the coaxial cable be lengthened as this 
may de-tune the circuitry and influence 
performance. 
 
Installation of the DT61 and DT63 small-pipe 
transducers follow the same procedures as the DT60 
and DT62 standard pipe type. The only difference is 
that the small pipe transducers utilize an integral 
clamping mechanism for pipe mounting and the 
standard pipe units use a stainless steel strap. 
 
Installations requiring intrinsic safety should refer to 
the Appendix drawings covering these applications. 
PART 2 - TRANSDUCER INSTALLATION
Acoustic Couplant 
Types 
Small Pipe 
Transducers 
Intrinsic Safety 
Installations 