Part 4 - configuration and operation – Dynasonics D300_301 Series Doppler Ultrasonic Flow Meter User Manual
Page 22

 
Rev. 2/00 -4.3- 300/301 
B. Totalizer Calibration — Series 301 
 
1. If an oscilloscope or signal counter is available, a 
signal from test point TP10 will show the totalizer 
count rate at a particular flow. Refer to the drawing in 
the Appendix titled 091-1048-102 Series 300 
Waveforms. The pulse rate at TP10 can be x 10, x 
100, x 1000 or x10,000 of totalizer count rate, at the 
CTR output on terminal strip. For example, if the 
pulse rate is 200 milliseconds (multiplier switch SW 1 
programmed for 100), then the counter rate at CTR 
terminals will be scaled to provide a count every 20 
seconds. The total counter output (CTR) range can 
be adjusted from approximately 0.5 to 5 seconds with 
a full scale flow indication. 
 
Totalizer Adjustment: 
 
(CTR CAL) - (pot R90 and DIP SW 1). The multiplier 
DIP switch SW 1 may be programmed as follows: 
 
SW 1 Position 
If an oscilloscope or signal counter is not available, 
then use the following procedure. The Pot R90 has a 
range of 0.5 to 5 seconds. SW1 is then used to 
increase this rate all the way up to 50 minutes. 
 
2. Stopwatch Procedure - with a known flow rate 
established on the flow meter digital display, time the 
count rate on the totalizer by a stopwatch, measuring 
the time between totalizer counts. If the totalizer is 
reading lower than desired, turn CTR, CAL pot R90 
PART 4 - CONFIGURATION AND OPERATION
SW1-1
SW1-2
MULTIPLIER
OFF
OFF
X 10
ON
OFF
X 100
OFF
ON
X 1000
ON
ON
X 10000