The following list outlines how to install the DT6
transducer for optimal performance, highest reliability
and greatest accuracy:

1. Select a transducer site at least 10 pipe diameters

downstream from bends, or fittings and 5 pipe
diameters upstream. A symmetrical flow pattern is
necessary for accuracy and repeatability over the

PART 2 - TRANSDUCER INSTALLATION

Table 2.1

1

1

The D300/301 system will provide repeatable measurements on piping systems that do not meet

these requirements, but the accuracy of these readings may be influenced to various degrees.