Part 4 - configuration and operation – Dynasonics D300_301 Series Doppler Ultrasonic Flow Meter User Manual
Page 21

 
Rev. 2/00 -4.2- 300/301 
FIELD CALIBRATION PROCEDURES
 
After the Equipment is properly installed and 
operating, there may be a need to field calibrate the 
flowmeter. This could be caused by a number of 
reasons: 
 
1. Using a size pipe other than the one specified 
when factory configured.
2. When used to measure a liquid with a sound
speed that is different than the one specified when 
factory configured. 
3. When proper lengths of straight pipe are not
available.
4. Percent of solids are greater than 2%.
5. Operating off turbulence or non-linear suspended
solids.
 
 
Field Calibration is as follows: 
 
A. To calibrate the flow rate indicators (meter) 
against a known flow: 
 
1. With an established flow, the digital readout and/
or current outputs on D300/301 can be adjusted in 
accordance with a verified flow measurement, by 
adjusting pot R30 (SPAN) adjustment. Turning 
the pot C.C.W. decreases the reading. This is a 
20 turn, stop-less potentiometer. (Note: Pot R60 
will adjust the 0-1 mA output only). 
2. It is desired to adjust 4-20 mA to other than
factory calibration, pot R50 can be used to adjust 
the 20 mA span. 
3. Adjust damping pot R110 for desired flow
response. CW rotation increases flow damping. 
The total flow damping range can be varied from 
about 5 seconds to 50 seconds. This setting is 
usually used to smooth out chart recorder 
readings. 
PART 4 - CONFIGURATION AND OPERATION
Field Calibration 
Procedure 