beautypg.com

Valve operation – AAON LC-054 User Manual

Page 34

background image

34

Procedure for Starting the Pump after

Servicing:

1. Ensure serviced pump is fully re-

assembled including all seal flush lines and

drain plugs.

2. Fill the dry casing with system fluid by

opening the seal flush line interconnecting

valve and the air vent fitting.

3. Allow the pressure to equalize in the two

casings, if necessary, by opening seal flush

line interconnected valve.

4. Unlock the discharge valve as per

instructions below.

5. Unlock the suction valve as per

instructions below.

NOTE: Keep hands and tools away from

locked suction valve arm, as the differential

pressure may cause the arm to rotate quickly

with force when unlocked.

6. Close the seal flush line interconnect

valve and restart pump.

Valve Operation

Refer to the valve illustrations on the

following pages.

Discharge Valve

This valve performs the dual function of

automatically sealing the discharge of the

inactive pump when one pump is running

and can manually be closed and locked to

isolate one pump for service.

Automatic Flapper Operation

In the flapper mode the two halves of the

discharge valve are free to pivot

independently under normal operating

conditions. The locking handle (3) should be

secured with the set screw (11) in the

vertical position with the center pin of the

locking arm (4) trapped by the locking

handle (3).

Manual Valve Locking:

The locking feature of this valve is to ensure

a positive seal (leak proof) of the discharge

port on the pump to be serviced.

Note: Ensure the pump to be isolated is not

operating before attempting to release the

locking mechanism. Failure to do so may

result in injury to the operator and/or

damage to the pump.

Locking

1. Loosen discharge side set screw (11) to

release the locking handle (3).

2. Rotate the discharge side locking handle

(3) so that the handle points toward the

pump to be serviced and secure in the

horizontal position, using set screw (11).

This releases the discharge locking arm (4).

3. Rotate discharge valve shaft (16) towards

the pump to be isolated. The orientation of

the shaft is indicated by the center pin on the

locking arm. (4).

4. Raise the locking handle (3) so that the

cam on the base of the handle forces the pin

of the locking arm (4) towards the pump to

be isolated. The locking handle (3) should

be raised to between 45 degrees and the

vertical position.

5. Tighten set screw (11) to lock the locking

handle (3) in position

THIS HANDLE SHOULD NOT BE

ROTATED PAST THE VERTICAL

POSITION.

Note: Ensure the isolated pump is not

operating before attempting to release the

locking mechanism. Failure to do so may

result in injury to the operator and/or

damage the pump.

This manual is related to the following products: