31 IOM-DPCDPR-A: FEB 2014
Example:
Determine the maximum vertical spacing of the seals for
the DPR-2200/-
10…10kPa transmitter when measuring
the density of liquid between 0.6 and 1.2g/cm³. The
sealing system uses AK-20 silicone oil with a density of
0.945g/cm³.