A

13 IOM-DPCDPR-A: FEB 2014

G1/4-type connector with G1/4 thread,
1/2"NPT -type connector with 1/2"NPT tread,
R1/2-type connector with R1/2 tread,
CG1/2-type connector with G1/2 tread and flush diaphragm,
other connection types by arrangement.

5.3. DPR-2000ALW, DPR-2200ALW Measurement ranges and metrological
parameters.

5.3.1. DPR-2000ALW, Measurement ranges

N

Nominal measuring range

(FSO)

Minimum set range

Rangeability

Overpressure

limit

Static pressure

limit

1

0...16bar

(0...1,6MPa)

1,6bar

(160kPa)

10:1

250, 320bar

(40bar for P-type connector)

2

0...2,5bar

(0...250kPa)

0,25bar (25kPa)

10:1

3

0...1bar

(0...100kPa)

70mbar (7kPa)

14:1

4

0...0,25bar

(0...25kPa)

10mbar (1kPa)

25:1

5

-100...100mbar (-10...10kPa)

4mbar

(0,4kPa)

50:1

6

-5...70mbar

(-0,5...7kPa)

4mbar

(0,4kPa)

18:1

7

-0,5...0,5bar

(-50...50kPa)

0,1bar

(10kPa)

10:1

8

-25...25mbar

(-2,5...2,5kPa) 2,5mbar (0,25kPa)

20:1

20 bar

5.3.2. DPR-2200ALW, Measurement ranges

Nominal range

(FSO)

Minimum set

range

Vertical spacing

of diaphragm

seals.

Maximum configurable range dependent

on the actual vertical spacing of

diaphragm seals. (m

)

Static

pressure

limit

-160...160 mbar

0,1 mH2O

1,7m

[1,6+( vertical spacing of sealsx94)]mH

2

O

40bar

-0,5...0,5 bar

0.5 mH2O

6m

[5+(vertical spacing of sealsx1,04)]mH

2

O

40bar

-1,6...2 bar

1,5 mH2O

15m

[20+(

vertical spacing of seals

x1,04)]mH

2

O

40bar

-1,6...16 bar

1bar

15m

16bar

40bar

The maximum vertical diaphragm seal spacing shown in the table applies to level measurement,
ensuring that it is possible to set the zero point of the transmitter when the tank is empty.
For measurements of density or phase boundaries (in the sugar and chemical industries and in
refineries) the vertical spacing of the diaphragm seals can be larger.

5.3.3. DPR-2000ALW, DPR-2200ALW Metrological

parameters

Accuracy

± 0,075% (FSO) for the DPR-2000ALW nominal range

Long term stability

accuracy / 3 years

Error due to supply voltage changes

± 0,002%(FSO)/1V

Thermal error

± 0,08%(FSO)/10ºC

Thermal error for the whole thermal
compensation range

± 0,3%(FSO)

Zero shift error for static pressure*

± 0,08 % (FSO)/10bar

± 0,01 % (FSO)/10bar (for range n°4)

± 0,03 % (FSO)/10bar (for range n°5, 6)

Cut-off on radical characteristic curve

up to10% of flow.

* zeroing in static pressure conditions with zero differential pressure eliminate this error.

5.3.4. DPR-2000ALW, Pressure Connectors

DPR-2000/ALW

– C-type connector to mount together with a valve manifold see fig.9,

DPR-2000/ALW with single direct diaphragm seal

– as in the example (fig.10) or with other diaphragm seals in

accordance with “DIAPHRAGM SEALS” catalogue cards.

5.3.5. DPR-2200ALW. Pressure Connectors - diaphragm seals

– see “DIAPHRAGM

SEALS

”catalogue cards.