B&K Precision 9116 - Programming Manual User Manual
Page 17
![background image](/manuals/320504/17/background.png)
Unit: V| mV| uV
Query syntax: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Return parameter:
[SOURce:]VOLTage:PROTection[:LEVel]
This command is used to set the software-voltage protection value.
Command syntax: [SOURce:] VOLTage:PROTection[:LEVel]
Parameter: MIN TO MAX|MIN|MAX|DEF
Parameters: V |mV| uV
Query syntax: [SOURce:] VOLTage:PROTection[:LEVel]?
Parameters: None
Return parameter:
[SOURce:]VOLTage:PROTection:DELay
This command is used to set the software-voltage protection delay time.
Command syntax: [SOURce:]VOLTage:PROTection:DELay
Parameter: 0.001~0.6
unit: s
Query syntax: [SOURce:]VOLTage:PROTection:DELay?
Parameter: 0.001~0.6
Return parameter:
[SOURce:]VOLTage:PROTection:STATe
This command is used to set the software-voltage protection state.
Command syntax: [SOURce:]VOLTage:PROTection:STATe
Parameter: 0|1|ON|OFF
Query syntax: [SOURce:]VOLTage:PROTection:STATe?
Parameter: None
Return parameter: 0|1
[SOURce:]VOLTage:PROTection:TRIGgered?
This command is used to query the executing state of over voltage protection. If “1”, this
indicates the OVP circuit has been triggered and the OVP state is not cleared. If “0”, the OVP
circuit is not triggered.
Command syntax: [SOURce:]VOLTage:PROTection:TRIGgered?
Return value: 0|1
PROTection:CLEar
This command is used to clear the OVP state. Before sending this command, please increase
the upper limitation of OVP or reeeduce the output voltage.
Note: Please remove the device that tripped this protection and then send the command again.
Command syntax: PROT:CLE
[SOURce:]VOLTage:LIMit[:LEVel]