[SOURce:]FALL[:LEVel]
This command is used to set the voltage falling time of the power supply.
Command syntax: [SOURce:] FALL [:LEVel]
Unit: s
parameter: 0~65.535
Query syntax: [SOURce:] FALL [:LEVel]?
Return parameter: 0~65.535
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
This command is used to set the output current value.
Command syntax: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
Parameter: MIN TO MAX|MIN|MAX|DEF
Unit: A |mA |uA
Query syntax: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Parameter: None
Return parameter:
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]{
This command is used to set a current value to be triggered. This value will be set after the
instrument receives a trigger signal. Sending CURRent command will not impact this
command’s current setting value. Sending a query command will return the original setting
value.
Command syntax: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
Parameter: MIN TO MAX|MIN|MAX|DEF
Unit: A
Query syntax: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
Return parameter:
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
This command is used to set the output voltage value.
Command syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
Parameters: MIN TO MAX|MIN|MAX|DEF
Unit: V| mV| uV
Query syntax: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Parameter: None
Return parameter:
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]{
This command is used to set a voltage value to be triggered. This value will be set when the
instrument receives a trigger signal. Sending VOLTage command will not impact this command’s
voltage setting value. Sending a query command will return the original setting value.
Command syntax: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Parameters: MIN TO MAX|MIN|MAX|DEF