8
Auxiliary Inputs
Trigger In
Impedance
1 kΩ or 50 Ω
Polarity
POS or NEG
Connector
BNC Front
Input Voltage Range
1 kΩ: ±10 V. 50 Ω: ±5 V
Threshold
Level
–5.0 V to 5.0 V
Resolution
0.1 V
Trigger Jitter
2.0 ns to 4.5 ns (Typical)
Trigger Mode
Minimum Pulse Width
20 ns
Trigger Hold-off
832* sampling_period – 100 ns
Delay to Analog Out
128* sampling_period + 250 ns
Gate Mode
Minimum Pulse Width
1024* sampling_period + 10 ns
Delay to Analog Out
640* sampling_period + 260 ns
Event Input
Impedance
1 kΩ or 50 Ω
Polarity
POS or NEG
Connector
BNC Front
Input Voltage Range
1 kΩ: ±10 V. 50 Ω: ±5 V
Threshold
–5.0 V to 5.0 V
Resolution
0.1 V
Sequence Mode Mode
Minimum Pulse Width
20 ns
Event Hold Off
1024* Sampling Period + 10 ns
Delay to Analog Out
640* Sampling Period + 280 ns (Jump timing: Asynchronous jump)
External Clock IN
Input Voltage Range
0.2 V
p-p
to 0.8 V
p-p
Impedance
50 Ω, AC coupled
Frequency Range
600 MHz to 1.2 GHz
Clock Divider
1/1, 1/2, 1/4…1/32
1/2, 1/4…1/32
Connector
BNC Rear
Reference Clock IN
Input Voltage Range
0.2 V
p-p
to 3.0 V
p-p
Impedance
50 Ω, AC coupled
Frequency Range
10 MHz, 20 MHz, 100 MHz (with ±0.1%)
Connector
BNC Rear
Phase Lock IN
Input Ranges
5 MHz to 600 MHz (acceptable frequency drift is ±0. 5%)
Input Voltage Range
0.2 V
p-p
to 3 V
p-p
Impedance
50 Ω, AC coupled
Multiple Rate
1 to 240
1 to 120
Connector
BNC Rear
Add IN
For each analog channel
Impedance
50 Ω, DC coupled
DC Gain
1
Bandwidth
DC to 100 MHz at –3 dB
Input Voltage Range
± 1.0 V
Connector
BNC Rear