beautypg.com

Characteristics, Digital phosphor oscilloscopes – Atec Tektronix-DPO4104 User Manual

Page 7

background image

Digital Phosphor Oscilloscopes

DPO4000 Series

Oscilloscope • www.tektronix.com/oscilloscopes

8

Characteristics

Vertical System

DPO4032

DPO4034

DPO4054

DPO4104

Input Channels

2

4

4

4

Analog Bandwidth (–3 dB)

350 MHz

350 MHz

500 MHz

1 GHz

5 mV/div to 1 V/div

Calculated Rise Time

1 ns

1 ns

700 ps

350 ps

5 mV/div (Typical)

Hardware Bandwidth Limits

20 MHz or 250 MHz

Input Coupling

AC, DC, GND

Input Impedance

1 MΩ ±1%, 50 Ω ±1%

Input Sensitivity, 1 MΩ

1 mV/div to 10 V/div

Input Sensitivity, 50 Ω

1 mV/div to 1 V/div

Vertical Resolution

8 bits

Max Input Voltage, 1 MΩ

250 V

RMS

with peaks ≤±400 V

Max Input Voltage, 50 Ω

5 V

RMS

with peaks ≤±20 V

DC Gain Accuracy

±1.5% with offset set to 0 V

Offset Range

1 mV/div to 50 mV/div ±1 V

50.5 mV/div to 99.5 mV/div ±0.5 V

100 mV/div to 500 mV/div ±10 V

505 mV/div to 995 mV/div ±5 V

1 V/div to 5 V/div ±100 V

5.05 V/div to 10 V/div ±50 V

Channel-to-Channel Isolation

≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth

(Any Two Channels at Equal
Vertical Scale)

Horizontal System

DPO4032

DPO4034

DPO4054

DPO4104

Maximum Sample Rate

2.5 GS/s

2.5 GS/s

2.5 GS/s

5 GS/s

(all channels)

Maximum Record Length

10 M points

(all channels)

Maximum Duration at Highest

4 ms

4 ms

4 ms

2 ms

Sample Rate (all channels)

Timebase Range (s/div)

1 ns to 1,000 s

400 ps to 1,000 s

Timebase Delay Time Range

–10 divisions to 50 s

Channel-to-channel Deskew Range

±100 ns

Long Term Sample Rate and

±5 ppm over any ≥1 ms interval

Delay Time Accuracy

Delta Time Measurement Accuracy

±(1/sample rate + 5 ppm x |Reading| + 0.4 ns)